71000-0 SETAFLASH ‘SERIES 7PLUS’ CLOSED CUP TESTER ASTM
D3278; ASTM D3828 - IP303 Part 1 and 2; ISO 3679; ISO 3680; BS 6664 Parts 3 and
4;BS 3900 Part A13; BS EN 456; ASTM E502; UN Class 3 Non-viscous
Flammable Liquids; CHIPS Regulations; Classification of Dangerous goods for
Carriage: Viscous and Non-Viscous liquids; EPA 1020 A and B; IP / ASTM Small
Scale draft method
·
Fully automatic operation ·
-30°C to 100°C temperature range ·
Flash point test in less than 2 minutes ·
Small sample size - 2ml ·
Rapid equilibrium and ramp modes ·
Programmable test parameters ·
Integral printer and RS232 interface
Setaflash
'Series 7Plus' is similar to the 'Series 7' instrument but incorporates
electronic peltier cells which enable testing of samples at sub-zero SUPPLIED
WITH: 2
ml syringe, -30° to 100°C secondary reference thermometer with certificate of
calibration and instruction manual. Note:
A supply of cooling water is required (for -30°C the flow must be 0.6litre/min
at 5°C or lower). Temperature
Range:
-30° to 100°C Test Modes:
Rapid Equilibrium and Ramp Test
Duration, RE Mode:
1 min. standard,
or user selectable 1 to 99 min. Test
Duration, Ramp Mode: Variable,
user definable up to 5.5°C/min. within 99 min. and100°C limits Computer
Interface:
RS232C Voltage
Range:
110/120V and
220/240V 50/60 Hz (switchable) Power:
200W
(maximum) Size
(HxWxD):
21 x 49 x 30cm Weight: 11.2kg ACCESSORIES 71001-0 SETA-VENT Portable fume extraction and draft shielding The
Seta-Vent is suitable for use with any of the Seta range of Flash Point Testers,
and is ideal for use where undesirable fumes need to be removed or where
protection from air currents is required. The unit incorporates fan and controller and can be
dismantled for easy storage when not in use. Note: Flexible ducting is not supplied.
|