Vapor Pressure Automatic Mini Methods 80000-0 SETAVAP ANALYZER (click on picture to enlarge) ASTM D5188; D5191; IP 394; IP 409; EN 13016-1; EN 13016-2 Correlates with: ASTM D323; D2533; D4953; D5482; D6378; IP 69; IP 402; ISO 3007;
The
80000-0 Setavap Analyzer is a vapor pressure tester with a test pressure range
of 0-600kPa and a temperature range from 25°C below ambient to 100°C. The
instrument is fully automatic, including the septum change, sample draining, and
print-out of test results. The instrument features six modes of operation,
including a Reid vapor pressure mode. In the vapor pressure modes, the
operator can select single or multiple vapor pressure tests on a sample at
static, ramping or stepped temperatures. The vapor-liquid mode enables
determination of the temperature which produces a pre-set pressure. Calibration
is automated and can be carried out at full vacuum, atmospheric pressure, or
high pressure. A calibration certificate can be printed out. SUPPLIED
WITH: 5ml
syringe, 3 needles, septum strips, paper/ink package, flask, mains lead and
instruction manual. Note: Seta 22422-6 Vacuum Pump is required for operation ACCESSORIES 22422-6
VACUUM PUMP
81000-2 SETAVAP 2 (click on picture to enlarge) ASTM D5191; IP 394; IP 409 (37.8°C); IP 481; EN 13016-1; EN
13016-2; Correlates with: D323; D4953; D5482; D6378; IP 69; IP 402; ISO3007; EN 12;
The
Setavap 2 is an automated instrument for the measurement of the vapor pressure
of gasoline, solvents, light crude oils, and other similar products. All
functions are accessed via a two button keypad and a rotary control. The digital
display leads the operator through the test sequence, displays the instrument
status and the test results. An audible prompt is sounded when an action is
required from the operator. A sample is injected through a septum into a fixed
volume chamber that is maintained at 37.8 °C ± 0.1 °C. The vapor to liquid
ratio is 4:1. The pressure is automatically measured at one minute intervals
until three readings are within 0.1 kPa, which determines the end of the test.
Results can be displayed as total pressure (Ptot), dry vapor pressure equivalent
(DVPE), EPA requirements, or Reid (crude) using pre-programmed correlation
equations, and may be printed via the integral RS232 port and optional 81002-0
printer. A facility is included to allow two part tests (Pabs) to be conducted
at 37.8 °C. All
valve operations are automated including sample draining and chamber venting and
purging. The valves utilise solvent resistant Kalrez seals Calibration of
temperature and pressure is automated and simple to carry out. SUPPLIED
WITH: septum
holder, waste container, mains lead, and instruction manual. Test
Temperature: 37.8 °C (100°F) ±0.1°C |