15400-2 SETA OXIDATION STABILITY CONSOLE BATHASTM D525-IP 40; ASTM D873-IP 138; BS 2000 Parts 40 and 138; EN 7536; ISO 7536; NF M07-012; NF M07-013; FTMS 791b 3352; FTMS 791b 3354; JIS K2287
Measures the induction period for oxygen in gasoline, and allows the weight of gum and residue in aviation fuel to be determined by direct weighing. The Seta Oxidation Stability Console Bath comprises a heated water bath with reflux condenser, a cold water plunge tank for cooling test vessels after each test, and an integral system for charging oxidation vessels prior to test. Includes oxygen charging lead (15620-0). Pressure can be monitored automatically using a Seta Auto Oxidation Control Unit (Seta 15450-2) or manually using either a Portable Cambridge Pressure Recorder (Seta 15611-0) or an oxygen pressure gauge (Seta 15600-0). The bath can accommodate up to four Oxidation Vessels. Capacity:
60 litres ACCESSORIES REQUIRED 15802-0 SETA
OXIDATION VESSEL, polished steel with hexagonal
top closure and ‘O’ ring seal. Complete
with stem, needle valve, quick connection
Schrader male coupling for oxygen charging,
and fitting for a transducer, pressure transmitting
capillary or pressure gauge. Fitted with
safety burst disc assembly and discharge tube.
Stainless steel burst disc ruptures within 10ms
at 1530kPa ±10%, fatigue life greater than 100,000
cycles up to 90% of max. pressure. THERMOMETERS: ASTM22C/IP24C or
ASTM22F/IP24F OPTIONAL 15650-0 WATER
PURIFIER, to minimize scale production. Suitable
for domestic water supply, 60L/hr max. flow.
Comprises disposable cartridge with capacity
for 960L soft water or 180L hard water. 15450-2 SETA AUTO OXIDATION CONTROL UNIT
An automatic
pressure monitoring system for use with 15400-2,
but can be used with other suitable baths. Comprises
two digital pressure monitors (each capable of
monitoring two vessels) with a membrane switch control
panel. The unit can also be used to display oxygen
pressure whilst charging vessels. The
control unit can be connected to a computer via an RS232
port, and the Seta OXS Software (supplied) used for
automatic end point detection, induction period determination,
data logging, and graphical/printed output
functions. The maximum run time may be selected
as either 24 or 96 hours. ACCESSORIES REQUIRED 15455-0 TRANSDUCER,
complete with lead (1 required for each oxidation vessel). Supplied with test
certificate. OPTIONAL 15620-0 SETA OXYGEN CHARGING LEAD, for connecting oxygen supply to oxidation vessel when not using 15400-2 bath. CALIBRATION, VERIFICATION & TRAINING 80005-0 PRESSURE CALIBRATOR, for recalibrating pressure monitors
For information on our solid block oxidation
baths please call 574/271-7020 or email us at sales@lazarsci.com
|